A Limit Test for Free Ferricyanide and Ferrocyanide in Sodium Nitroprusside
-
-
Abstract
A rapid method of limit test for free ferricyanide and ferrocyanide in sodiunm nitroprussideby column chromatography combined with spectrophotometric method is presented.The rnethod is based on the fact that ferricyanide has a maximum absorption at thewave length of 420nm while ferrocyanide hasn't unless it is oxidized with 2N HNO_3to ferricyanide.
-
-