GAS CHROMATOGRAPHY OF VANILLIN IN ANTI-EPILEPSY TABLETS
-
Graphical Abstract
-
Abstract
The method was used for determination of vanillin content in an anti-epilepsy tablets. The average recovery was 99.37±1.1% (n=5). The anal. conditions were:5% polyethylene glycol on 101 white support in 2m * 3mm column at 135℃;Flame ionization detector, H2 48 ml/min, air 430 ml/ min; detector and injection port at 190℃; sample size 0.5 μl.
-
-